REHATRON™alpha - technical functioning

Rehatron - Technik

Technically speaking, REHATRON™alpha is a generator that generates magnetic impulses with a high intensity and possesses a unique impulse character. The extremely powerful energy impulses are short and can reach intracellular structures. At intervals of nano or micro seconds the electromagnetic impulses can activate cells, energize cell molecules and make them maintain or achieve their original oscillation.

REHATRON™alpha consists of a power supply unit with a big capacitor that can be charged to 30,000 V (Volt). The electrical energy stored in the capacitor is discharged by means of the thyristor switch in the area of nano and micro seconds and transferred to the organism via the unique treatment loop.

The capacity of the high-performance capacitor and the inductivity of the treatment loop consisting of three turns form an oscillatory circuit so that the electrical energy of the capacitor vibrates approx. 25 times along a damped curve according to the ohmic losses of the oscillatory circuit. The slew rate of the first vibrations lies in the area of nano seconds. The maximum magnetic performance leads to the necessary potential in the organism. The magnetic field has a force of approx. 6,000 to 12,600 Ampere coils (6 – 12.6 kAw) in its peak value. The generated electricity increases the low cell membrane potential in case of diseases and injuries.

REHATRON™alpha is produced according to VDE and the European guidelines for medical products and was approved technically and medically for the first time in 2005 within the purview of „non-invasive induction therapy“. All involved companies are EN ISO 13485-certified.

REHATRON™alpha - Technical data

Power supply
supply voltage: 230 V ~
supply frequency: 50 - 60 Hz
connected value: 700 VA

 

Pulse shaping
charging unit: 500 J/s
potential adjustable (intensity): 15 - 30 kV
capacitor: 0,11 µF
magnetic field coil: approx. 4,75 µH
discharge, peak value: 2 - 4kA
resonant base frequency: 200 - 220 kHz
pulse duration: 100 - 150 µs
flow density in the middle of the coil, peak value: 40-80 mT
repetition rate adjustable (Hz): 1-3 Hz
duration of treatment adjustable: 5-20 min.

 

Isolation of the magnetic field coil
nominal voltage of the isolation: 30 kV
test voltage of the isolation: 65 kV DC 1 Minute

 

Contact

Novotech Elektronik GmbH
Freistädter Str. 38
A-4210 Gallneukirchen


E-Mail:    This email address is being protected from spambots. You need JavaScript enabled to view it.
E-Mail:    This email address is being protected from spambots. You need JavaScript enabled to view it.

Anwender-Support
Informationen für Patienten
Kooperationen
V.E.I.T. e.V.